• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Medical Devices

  • Print
  • Share
  • E-mail

IDE Early/Expanded Access

Introduction

An unapproved medical device may normally only be used on human subjects through an approved clinical study in which the subjects meet certain criteria and the device is only used in accordance with the approved protocol by a clinical investigator participating in the clinical trial. However, there may be circumstances under which a health care provider may wish to use an unapproved device to save the life of a patient or to help a patient suffering from a serious disease or condition for which there no other alternative therapy exists. Patients/physicians faced with these circumstances may have access to investigational devices under one of four main mechanisms by which FDA may make an unapproved device available:

These mechanisms can be utilized during a certain time-frame in the IDE process if the criteria are met. FDA approval is required except in the case of emergency use. The mechanisms are summarized below followed by an in depth discussion of criteria and requirements.

This graph compares the Traditional IDE Study with the various timeframes in the device development process in which early access provisions can be utilized. The Traditional IDE Study is conducted after IDE approval until completion of the IDE approved study. Emergency Use can be utilized throughout the device development process, that is, before or after IDE approval up to marketing approval. Single Patient/Small Group Access (or Compassionate Use) can be utilized after IDE approval until marketing approval. Treatment Use can be used during the IDE study after data suggests that the device is effective until marketing approval. Continued Access can be utilized after the IDE approved study is completed until the device receives marketing approval.

Emergency Use

Emergency situations may arise in which there will be a need to use an investigational device in a manner inconsistent with the approved investigational plan or by a physician who is not part of the clinical study. Emergency use of an unapproved device may occur before an IDE is approved.

Criteria:
  • Life-threatening or serious disease or condition
  • No alternative
  • No time to obtain FDA approval
Time-frame:Before or after initiation of the clinical trial


Emergency Research

There are special cases under emergency research in which the human subject is in a life-threatening situation and it is not feasible to obtain informed consent. In order to allow such research to proceed, special provisions for exception from informed consent requirements must be met. In addition, the IRB and a physician not participating in the investigation must review and approve the investigation. The sponsor must also submit a separate IDE application to FDA.

Detailed information on conditions for exception from informed consent can be found under Informed Consents, Exception from Informed Consent Requirements for Emergency Research.

Compassionate Use(or Single Patient/Small Group Access)

The compassionate use provision allows access for patients who do not meet the requirements for inclusion in the clinical investigation but for whom the treating physician believes the device may provide a benefit in treating and/or diagnosing their disease or condition. This provision is typically approved for individual patients but may be approved to treat a small group.

Criteria:
  • Serious disease or condition
  • No alternative
Time-frame: During clinical trial

FDA recognizes that there are circumstances in which an investigational device is the only option available for a patient faced with a serious, albeit not life-threatening, disease or condition. In these circumstances, FDA uses its regulatory discretion in determining whether such use of an investigational device should occur.

Prior FDA approval is needed before compassionate use occurs. In order to obtain Agency approval, the sponsor should submit an IDE supplement requesting approval for a protocol deviation under section §812.35(a) in order to treat the patient. The IDE supplement should include:

  • A description of the patient's condition and the circumstances necessitating treatment
  • A discussion of why alternatives therapies are unsatisfactory and why the probable risk of using the investigational device is no greater than the probable risk from the disease or condition
  • An identification of any deviations in the approved clinical protocol that may be needed in order to treat the patient
  • The patient protection measures that will be followed. (Informed consent, concurrence of IRB chairperson, clearance from the institution, independent assessment from uninvolved physician, authorization from IDE sponsor)

The physician should not treat the patient identified in the supplement until FDA approves use of the device under the proposed circumstances. In reviewing this type of request, FDA will consider the above information as well as whether the preliminary evidence of safety and effectiveness justifies such use and whether such use would interfere with the conduct of a clinical trial to support marketing approval.

If the request is approved, the attending physician should devise an appropriate schedule for monitoring the patient, taking into consideration the investigational nature of the device and the specific needs of the patient. The patient should be monitored to detect any possible problems arising from the use of the device. Following the compassionate use of the device, a follow-up report should be submitted to FDA as an IDE supplement in which summary information regarding patient outcome is presented. If any problems occurred as a result of device use, these should be discussed in the supplement and reported to the reviewing IRB as soon as possible.

The above compassionate use criteria and procedures can also be applied when a physician wishes to treat a few patients rather than an individual patient suffering from a serious disease or condition for which no alternative therapy adequately meets their medical need. In this case, the physician should request access to the investigational device through the IDE sponsor. The sponsor should submit an IDE supplement that includes the information identified above and indicates the number of patients to be treated. Such a supplement should include the protocol to be followed or identify deviations from the approved clinical protocol. As with single patient compassionate use, a monitoring schedule should be designed to meet the needs of the patients while recognizing the investigational nature of the device. Follow-up information on the use of the device should be submitted in an IDE supplement after all compassionate use patients have been treated.

Treatment Use

An approved IDE specifies the maximum number of clinical sites and the maximum number of human subjects that may be enrolled in the study. During the course of the clinical trial, if the data suggests that the device is effective, then the trial may be expanded to include additional patients with life-threatening or serious diseases.

Criteria:
  • Life-threatening or serious disease
  • No alternative
  • Controlled clinical trial
  • Sponsor pursuing marketing approval
Time-frame:During clinical trial

A device that is not approved for marketing may be under clinical investigation for a serious or immediately life- threatening disease or condition in patients for whom no comparable or satisfactory alternative device or other therapy is available. During the clinical trial or prior to final action on the marketing application, it may be appropriate to use the device in the treatment of patients not in the trial under the provisions of the treatment investigational device exemptions (IDE) regulation. (§812.36)

The treatment use provision of the IDE facilitates the availability of promising new devices to desperately ill patients as early in the device development process as possible, before general marketing begins, and to obtain additional data on the device's safety and effectiveness. In the case of a serious disease, a device ordinarily may be made available for treatment use under this section after all clinical trials have been completed. In the case of an immediately life-threatening disease, a device may be made available for treatment use under this section prior to the completion of all clinical trials.

An "immediately life-threatening" disease means a stage of a disease in which there is a reasonable likelihood that death will occur within a matter of months or in which premature death is likely without early treatment. "Treatment use" of a device includes the use of a device for diagnostic purposes.

FDA would consider the use of an investigational device under a treatment IDE if:

  • The device is intended to treat or diagnose a serious or immediately life-threatening disease or condition.
  • There is no comparable or satisfactory alternative device or other therapy available to treat or diagnose that stage of the disease or condition in the intended patient population.
  • The device is under investigation in a controlled clinical trial for the same use under an approved IDE, or such clinical trials have been completed.
  • The sponsor of the investigation is actively pursuing marketing approval/clearance of the investigational device with due diligence.

    Applications for treatment use

    A treatment IDE application must include, in the following order:

    1. The name, address, and telephone number of the sponsor of the treatment IDE;
    2. The intended use of the device, the criteria for patient selection, and a written protocol describing the treatment use;
    3. An explanation of the rationale for use of the device, including, as appropriate, either a list of the available regimens that ordinarily should be tried before using the investigational device or an explanation of why the use of the investigational device is preferable to the use of available marketed treatments;
    4. A description of clinical procedures, laboratory tests, or other measures that will be used to evaluate the effects of the device and to minimize risk;
    5. Written procedures for monitoring the treatment use and the name and address of the monitor;
    6. Instructions for use for the device and all other labeling as required under section §812.5(a) and (b);
    7. Information that is relevant to the safety and effectiveness of the device for the intended treatment use. Information from other IDE's may be incorporated by reference to support the treatment use;
    8. A statement of the sponsor's commitment to meet all applicable responsibilities under the IDE regulations (21 CFR 812) and Institutional Review Boards regulations (21 CFR 56) and to ensure compliance of all participating investigators with the informed consent requirements of 21 CFR 50;
    9. An example of the agreement to be signed by all investigators participating in the treatment IDE and certification that no investigator will be added to the treatment IDE before the agreement is signed; and
    10. If the device is to be sold, the price to be charged and a statement indicating that the price is based on manufacturing and handling costs only.

    A licensed practitioner who receives an investigational device for treatment use under a treatment IDE is an "investigator" under the IDE and is responsible for meeting all applicable investigator responsibilities under 21 CFR 812, 21 CFR 50, and 21 CFR 56.

    Applications should be identified on the outside envelope as a treatment IDE application and reference the IDE number. The original and two copies should be mailed to the following address:

    Food and Drug Administration
    Center for Devices and Radiological Health
    Document Mail Center - WO66-G609
    10903 New Hampshire Avenue
    Silver Spring, MD 20993-0002

    FDA action on treatment IDE applications

    Approval of treatment IDE's

    Treatment use may begin 30 days after FDA receives the treatment IDE submission. FDA may notify the sponsor in writing earlier than the 30 days that the treatment use may or may not begin. FDA may approve the treatment use as proposed or approve it with modifications.

    Disapproval or withdrawal of approval of treatment IDE's

    FDA may disapprove or withdraw approval of a treatment IDE if:

    • The required criteria [§812.36(b)] are not met or the treatment IDE application does not contain the required information [§812.36(c)];
    • FDA determines that any of the grounds for disapproval or withdrawal of approval apply [§812.30(b)(1) through (b)(5)].  See Approval Process, FDA Actions, for additional information;
    • The device is intended for a serious disease or condition and there is insufficient evidence of safety and effectiveness to support such use;
    • The device is intended for an immediately life-threatening disease or condition and the available scientific evidence, taken as a whole, fails to provide a reasonable basis for concluding that the device:
      • may be effective for its intended use in its intended population; or
      • would not expose the patients to whom the device is to be administered to an unreasonable and significant additional risk of illness or injury;
    • There is reasonable evidence that the treatment use is impeding enrollment in, or otherwise interfering with the conduct or completion of, a controlled investigation of the same or another investigational device;
    • The device has received marketing approval/clearance or a comparable device or therapy becomes available to treat or diagnose the same indication in the same patient population for which the investigational device is being used;
    • The sponsor of the controlled clinical trial is not pursuing marketing approval/clearance with due diligence;
    • Approval of the IDE for the controlled clinical investigation of the device has been withdrawn; or
    • The clinical investigator(s) named in the treatment IDE are not qualified by reason of their scientific training and/or experience to use the investigational device for the intended treatment use.

    Notice of disapproval or withdrawal

    If FDA disapproves or proposes to withdraw approval of a treatment IDE, FDA will follow the procedures set forth in the IDE regulations [§812.30(c)] For more information, see  Approval Process, FDA Actions.

    Safeguards

    Treatment use of an investigational device is conditioned upon the sponsor and investigators complying with the safeguards of the IDE process and the regulations governing informed consent (21 CFR 50) and institutional review boards (21 CFR 56).

    Reporting requirements

    The sponsor of a treatment IDE must submit progress reports on a semi-annual basis to all reviewing IRB's and FDA until the filing of a marketing application. The date of these reports are based on the period of time since initial approval of the treatment IDE. After filing of a marketing application, progress reports must be submitted annually in accordance with the IDE regulation.

    See "Suggested Format For IDE Progress Report" under Reports for guidance on the content of a progress report. The progress report must also include the number of patients treated with the device under the treatment IDE, the names of the investigators participating in the treatment IDE, and a brief description of the sponsor's efforts to pursue marketing approval/ clearance of the device.

    The sponsor of a treatment IDE is responsible for submitting all other reports required under §812.150 (Reports), such as unanticipated adverse device effects and final reports. The reports are submitted as supplements to the original IDE application. See Reports for additional guidance.

    Continued Access

    FDA may allow continued enrollment of subjects after the controlled clinical trial under an IDE has been completed in order to allow access to the investigational medical device while the marketing application is being prepared by the sponsor or reviewed by FDA.

    Criteria:
    • Public health need or
    • Preliminary evidence that the devcie will be effective and there are no significant safety concerns
    Time-frame: After completion of the clinical trial

    The sponsor of a clinical investigation is permitted to continue to enroll subjects while a marketing application is being prepared by the sponsor and/or reviewed by the Agency if there is:

    • A public health need for the device; or
    • Preliminary evidence that the device is likely to be effective and no significant safety concerns have been identified for the proposed indication.

    The continued enrollment of subjects in an investigation while a marketing application is being prepared by the sponsor and/or reviewed by ODE is known as an "extended investigation." Extended investigations permit patients and/or physicians continued access to the devices while also allowing the collection of additional safety and effectiveness data to support the marketing application or to address new questions regarding the investigational device. The Continued Access Policy may be applied to any clinical investigation that meets the criteria identified above; however, it is intended to be applied late in the device development process, i.e., after the controlled clinical trial has been completed.

    A sponsor's request for an extended investigation should be submitted as an IDE supplement and include the following information:

    • A justification for the extension;
    • A summary of the preliminary safety and effectiveness data generated under the IDE;
    • A brief discussion of the risks posed by the device;
    • The proposed rate of continued enrollment (the number of sites and subjects);
    • The clinical protocol, if different from that used for the controlled clinical trial, as well as the proposed objectives for the extended study; and
    • A brief discussion of the sponsor's progress in obtaining marketing approval/clearance for the device.

    There is significant overlap between the treatment IDE regulation and the Continued Access Policy. Both the Continued Access Policy and the treatment IDE regulation are intended to provide additional access to an unapproved device, once preliminary evidence regarding safety and effectiveness is available to FDA. However, because a treatment IDE can be submitted earlier in the IDE process, i.e., once promising evidence of safety and effectiveness has been collected under the IDE but while the clinical study is ongoing, it could provide access to a wider group of patients at an earlier stage in the IDE process. The treatment IDE regulation also has a more narrow application than the Continued Access Policy in that treatment use is intended to address only those patients who have an immediately life-threatening or serious disease or condition whereas the Continued Access Policy, which is applied after completion of the clinical trial, may be considered for any clinical investigation.

-
-